TG-J216A Infrared Analyzer

 TG-J216A Infrared Analyzer is a kind of NRIR (nondispersive infrared) analyzer. The basic working princple is the selective absorption of gases with infrared ray. We use singal light source, the singal tube separated into two halvies as gas chamber and the advanced serial membrane micro sound detector as detector in this analyzer, and the electrical part of this analyzer is the embedded control system which is independently developed by our company. Comparing with the similar products, this analyzer has many advantages, such as high stability, high reliability and high selectivity and so on.

Application Range:


TG-J216A Infrared Analyzer is used for continuous measurment and analysis of CO, CO2, CH4, SO2, NH3, and NOx and so on. It is widly used in many fields, such as petroleum, metallurgy, fertilizer, chemical industry, environment protection, agriculture, medical and scientific research and so on.

Typitcal Application Range:


l        TG-J216A Infrared Analyzer is typitcally used in pollution sources monitoring, the continuous emission monitoring and the air qulity monitoring;

l        This analyzer is greatly used for detection of combustion efficiency.

Instrument Characteristics:


l        Using imported infrared sensors from German company MAIHAK, with a long life expectancy, high precision, fast response characteristics;

l        The flow controled by electronic flow controllor, avoid blockage;

l        Precision filter inside the instrument extending sensor lifetime greatly;

l        The whole english operation menu, easy to operate;

l        320×240 dot-matrix LCD, display the test gas concentration, date, time and temperature of the inside instrument at the same time;

l        Automatic measurement data storage, record time interval length can be set freely. Data can be display with curve or list, easy to check;

l        Upper and lower control point in the whole range area can be set up at will;

l        Standard RS232 or RS485 communication interface, both-way communication with computer.

Technical Specifications:

Test Gas

Min Measure Range ppm

Max Measure Range %

CO

0~10

0~100

CO2

0~10

0~100

CH4

0~100

0~100

C2H6

0~500

0~100

C2H4

0~1000

0~100

C2H2

0~500

0~100

C3H8

0~500

0~100

C3H6

0~500

0~100

C3H4

0~500

0~100

C4H10

0~500

0~100

C4H6

0~500

0~100

C6H14

0~500

0~100

CH

0~500

0~100

SO2

0~300

0~100

NH3

0~500

0~100

NO

0~300

0~100

N2O

0~50

0~100

SF6

0~300

0~100

Freon

0~500

0~100

C2H5OH

0~3000

0~100

CC14

0~5000

0~100

CHCL3

0~5000

0~100

CH3CL

0~500

0~100

l        Accuracy:                           ≤ ±1%F.S;

l        Stability:                      Zero point drift ≤ ±1%F.S/7d;

Range drift    ≤ ±1%F.S/7d;

l        Sample gas flow:                 500 ± 10ml/min;

l        Sample gas pressure:          inlet pressure ≤ 0.4MPa;

l        Gas-chamber Pressure:       ≤ 20KPa;

l        Response time:                   T90 ≤ 20 sec;

l        Pre-warm time:                   ≤ 30 min;

l        Output signal:               4~20mA or 0~10mA DC optional;

l        Contact Capacity:        120VAC 1A, 24VDC 1A;

l        Working power:                  220VAC ± 10%, 50Hz ± 5%;

l        Working environment:  Temperature: -5íŠ ~ +45íŠ;

Humidity:              ≤90%RH;

l        External dimension:             490mm (W) × 150mm (H) × 280mm (D);

l        Installation dimension:  442mm (W) × 136mm (H);

l        Weight:                         About 13.0 Kg;